![Fluke MP Series Скачать руководство пользователя страница 20](http://html.mh-extra.com/html/fluke/mp-series/mp-series_abridged-user-manual_2309932020.webp)
MP
Series
Users Manual, Rev. 1.0, Apr 2021
20
2.3 Electrical Specifications
Power
24 VDC ± 25%, 1 A
Ethernet
Connection
M12 socket, 4 wires (full duplex)
100 MBit/s, 100BASE-TX / IEEE 802.3u, Auto-Negotiation
electrically isolated from power supply
Addressing
fixed IP address
BootP (using UDP sockets 67 and 68)
Protocol
TCP/IP version 4
Default Settings
IP Address: 192.168.42.30, port 2727
Subnet Mask: 255.255.255.0
Gateway: 192.168.42.1
BootP: enabled per factory default
MAC address: see sticker on the scanner
RS485
Connection
full-duplex, not addressable, electrically isolated
9.6, 57.6, 115.2, 230.4 kBaud
Outputs
Analog
3 active current outputs, each adjustable 0/4 ... 20 mA, maximum load: 500
Ω,
minimal load: 47
Ω
common ground connection for all current outputs, electrically isolated to the GND
ground
Alarm
potential-free relay contacts: 30 V / 1 A
normally open / normally closed
Exposure Time
4
1
𝑓
𝑠𝑐𝑎𝑛
Response Time
5
20 ms +
1
𝑓
𝑠𝑐𝑎𝑛
Inputs
Trigger
+ 5 to 24 VDC pulse, high/low active
Functional Input
max. + 5 VDC
function depends on the specific system
Signal Processing
Stand-alone scanner
up to 3 sectors / zones, PC independent at runtime
Max, Min, Average, Peak/Valley Hold, Alarm setpoints
further signal processing function configurable through software
I/O module support
up to 10 sectors / zones, PC independent at runtime
Max, Min, Average, Peak/Valley Hold, Alarm setpoints
further signal processing function configurable through software
4
The exposure time is the minimum time during which the measured object must be present. The output value of the sensor is delayed.
5
Time interval between the instant of an abrupt change for the object temperature and its availability on the output
Содержание MP Series
Страница 16: ...MP Series Users Manual Rev 1 0 Apr 2021 16 1024 pixel 80 Hz scan rate Scan Angle FOV All models 90 ...
Страница 18: ...MP Series Users Manual Rev 1 0 Apr 2021 18 2 2 1 2 Far Focus Figure 2 2 Optical Diagrams Far Focus ...
Страница 37: ...Installation Ethernet 37 5 5 Close the dialog box by pressing on Save ...
Страница 44: ...MP Series Users Manual Rev 1 0 Apr 2021 44 7 2 Mounting Plate A MP MP Figure 7 1 Mounting Plate for Tripod ...