Perfomance Testing and Calibration
Ohms/nS Calibration
4
4-27
AC Volts Calibration
4-30.
1.
Rotate the selector knob to
(VAC). Remove the ohms connection and connect the
5700A to the
Ë
and
â
inputs on the GMM.
2.
Step 24:
Apply 1000.0V, 1 kHz. Allow for settling; then press
t
[Proceed].
3.
Step 25:
Apply 100.00V, 1 kHz. Allow for settling; then press
t
[Proceed].
4.
Step 26:
Apply 290.00V, 1 kHz. Allow for settling; then press
t
[Proceed].
5.
Step 27:
Apply 29.000V, 1 kHz. Allow for settling; then press
t
[Proceed].
6.
Step 28:
Apply 2.9000V, 1 kHz. Allow for settling; then press
t
[Proceed].
7.
Step 29:
Apply 290.00 mV, 1 kHz. Allow for settling; then press
t
[Proceed].
8.
Step 30:
Apply 29.00 mV, 1 kHz. Allow for settling; then press
t
[Proceed].
9.
Set the 5700A to Standby.
AC mA
µµµµ
A Calibration
4-31.
1.
Rotate the selector knob to
(mA
µ
A).
2.
Remove the ac connection and connect the 5700A to the
and
â
inputs
on the GMM.
3.
Step 31:
Apply 290.00 mA, 1 kHz. Allow for settling; then press
t
[Proceed].
4.
Step 32:
Apply 29.000 mA, 1 kHz. Allow for settling; then press
t
[Proceed].
5.
Step 33:
Apply 2.900 mA, 1 kHz. Allow for settling; then press
t
[Proceed].
6.
Step 34:
Apply 290.00
µ
A, 1 kHz. Allow for settling; then press
t
[Proceed].
7.
Step 35:
Apply 29.00
µ
A, 1 kHz. Allow for settling; then press
t
[Proceed].
AC Amps Calibration
4-32.
1.
Place the 5700A in Standby.
2.
Remove the
connection and connect the 5725A to the
á
and
â
inputs.
3.
Rotate the selector to
(Amps).
4.
Step 36:
Apply 10.00A, 1 kHz. Allow for settling; then press
t
[Proceed].
5.
Step 37:
Apply 1.0000A, 1 kHz. Allow for settling; then press
t
[Proceed].
6.
Step 38:
Apply 2.9000A, 1 kHz. Allow for settling; then press
t
[Proceed].
7.
Step 39:
Apply 0.2900A, 1 kHz. Allow for settling; then press
t
[Proceed].
Содержание 863
Страница 49: ...Maintenance Disassembly 3 3 5 OFF 1 6 Places 4 5 5 3 3 2 os5f eps Figure 3 1 Disassembly ...
Страница 50: ...867B 863 Service Manual 3 6 os6f eps Figure 3 1 Disassembly cont ...
Страница 52: ...867B 863 Service Manual 3 8 5 1 6 2 3 4 os30f eps Figure 3 2 Reassembly ...
Страница 94: ...List of Replaceable Parts Parts 5 5 15 FLUKE 86X 4001 Sheet 1 of 2 os31_1f eps Figure 5 2 A1 Main PCA ...
Страница 95: ...867B 863 Service Manual 5 16 FLUKE 86X 4001 Sheet 2 of 2 os31_2f eps Figure 5 2 A1 Main PCA cont ...
Страница 96: ...6 1 Chapter 6 Schematic Diagrams Title Page 6 1 A1 Main PCA Assembly 6 3 ...
Страница 98: ...867B 863 Service Manual 6 4 FLUKE 867 1001 Sheet 2 of 5 Figure 6 1 A1 Main PCA cont ...
Страница 99: ...Schematic Diagrams 6 6 5 FLUKE 867 1001 Sheet 3 of 5 Figure 6 1 A1 Main PCA cont ...
Страница 100: ...867B 863 Service Manual 6 6 FLUKE 867 1001 Sheet 4 of 5 Figure 6 1 A1 Main PCA cont ...
Страница 101: ...Schematic Diagrams 6 6 7 FLUKE 867 1001 Sheet 5 of 5 Figure 6 1 A1 Main PCA cont ...