Perfomance Testing and Calibration
Calibration
4
4-21
Glitch Capture Test
4-18.
1.
Turn on the GMM to mV.
2.
Press
,
vc
[View],
w
[Glitch Capture],
t
[Arm].
3.
Apply a .070V p-p Pulse 10 kHz Pulse wave.
4.
Check for the Glitch Capture PASS/FAIL indications shown in Table 4-21.
Table 4-21. Glitch Capture Performance Test
Input
Input Source
PASS
FAIL
.07V p-p Pulse, 10 kHz, 50% duty
PM5139 to the GMM volts input
Trigger occurs
No Trigger
External Trigger Test
4-19.
1.
Turn on the GMM to mV DC.
2.
Press:
,
v
[View],
u
[Trigger],
w
[Internal/External](to select
external),
x
[Exit].
3.
Press
v
[Single Shot] ,
t
[Arm].
4.
Apply a 2V p-p 100 kHz Pulse wave.
5.
Check for the PASS/FAIL indications shown in Table 4-22.
Table 4-22. External Trigger Performance Test
Input
Input Source
PASS
FAIL
2.0V p-p, 100 kHz Pulse, 50% duty
PM5139 to the GMM EXT TRIG input. Trigger occurs No Trigger
Содержание 863
Страница 49: ...Maintenance Disassembly 3 3 5 OFF 1 6 Places 4 5 5 3 3 2 os5f eps Figure 3 1 Disassembly ...
Страница 50: ...867B 863 Service Manual 3 6 os6f eps Figure 3 1 Disassembly cont ...
Страница 52: ...867B 863 Service Manual 3 8 5 1 6 2 3 4 os30f eps Figure 3 2 Reassembly ...
Страница 94: ...List of Replaceable Parts Parts 5 5 15 FLUKE 86X 4001 Sheet 1 of 2 os31_1f eps Figure 5 2 A1 Main PCA ...
Страница 95: ...867B 863 Service Manual 5 16 FLUKE 86X 4001 Sheet 2 of 2 os31_2f eps Figure 5 2 A1 Main PCA cont ...
Страница 96: ...6 1 Chapter 6 Schematic Diagrams Title Page 6 1 A1 Main PCA Assembly 6 3 ...
Страница 98: ...867B 863 Service Manual 6 4 FLUKE 867 1001 Sheet 2 of 5 Figure 6 1 A1 Main PCA cont ...
Страница 99: ...Schematic Diagrams 6 6 5 FLUKE 867 1001 Sheet 3 of 5 Figure 6 1 A1 Main PCA cont ...
Страница 100: ...867B 863 Service Manual 6 6 FLUKE 867 1001 Sheet 4 of 5 Figure 6 1 A1 Main PCA cont ...
Страница 101: ...Schematic Diagrams 6 6 7 FLUKE 867 1001 Sheet 5 of 5 Figure 6 1 A1 Main PCA cont ...