Perfomance Testing and Calibration
Performance Tests
4
4-7
Diode Test
4-7.
Rotate the GMM selector to Diode Test (
ä
). Connect the GMM and 5700A in
Configuration 2 (Figure 4-2,
ä
selected). Then perform the three steps called for in
Table 4-4.
Before proceeding to the next test, place the 5700A in Standby.
Table 4-4. Diode Test Performance Test
Step
GMM Range
Input
Input Source
Lower Limit
Upper Limit
1
Auto
2.9V
5700A
2.3
3.5
2
-2.9V
5700A
-2.3
-3.5
3
Manual
2.5000V
5700A
2.4986
2.5015