Fluke 6500
Users Manual
18
Table 3. Factory Programmed Auto-Tests for Class I Appliances
Tests
131 132 133 134 135 136
Visual
Inspection
Yes Yes Yes Yes Yes Yes
Earth Bond 25A (
Ω
)
0.10 0.10 No No 0.10 No
Earth Bond 200mA
No
No
0.10
0.10
No
0.10
Insulation (M
Ω
) 1.00
1.00
1.00
1.00 1.00 1.00
Touch
Current
(mA)
No No No No No No
Substitute
Leakage
(mA)
No No No No No No
Load/Leakage (VA/mA)
3000/3.5
3000/0.75 3000/3.5 3000/0.75 No
No
Table 4. Factory Programmed Auto-Tests for Class II Appliances
Tests
231 232 233 234
Visual Inspection
Yes
Yes
Yes
Yes
Earth Bond 25A (
Ω
) No No No No
Earth Bond 200mA
No
No
No
No
Insulation (M
Ω
) 2.00
2.00
2.00
2.00
Touch Current (mA)
0.25
0.25
No
No
Substitute Leakage (mA)
No
No
No
No
Load/Leakage (VA/mA)
3000/0.25 No
No
3000/0.25
Note: Test numbers 137 to 142 and 235 to 240 are reserved for future factory programmed tests.