Users Manual
Description of the Tests
21
Figure 4. Bond Test Connections
The display can show the following specific information:
LMT
>0.1
Ω
+LEAD
R
PE
may have exceeded the
recommended limit, possibly because
of the length of the lead.
> 19.99
Ω
R
PE
overrange.
Ø
Bond test lead has been zeroed.
Note:
In the case that an Earth bond test fails, the
pass/fail limit can be recalculated by using the
SETUP key and entering the length and cross
section of the mains lead.
Insulation Test (RISO)
Warning
•
The test voltage is 500V dc. Do not touch
the appliance during the insulation test!
If the test fails any metal parts of the
appliance could become live!
•
Always make sure that the test has
completed before disconnecting the
appliance leads to ensure that all
capacitances have discharged.
Caution
Do not perform the Insulation test on
appliances that failed the bond test or the
visual inspection test.
The test checks the resistance of the insulation between
•
the earth pin of the appliance cable plug (Class I)
or
•
the test probe to be applied to the appliance (Class II)
and the Live and Neutral pins of the appliance (pins are
connected together within the tester for this test).
The insulation test will be inhibited if the tester detects a
terminal voltage >30 Vrms prior to initiation of the test.