RPM4™ OPERATION AND MAINTENANCE MANUAL
Page
88
4.4.4
PROGRAM MESSAGE DESCRIPTIONS
#
Purpose
To allow the host PC to communicate with a device connected to the RPM4 COM2 port.
Classic
“#xx”
Arguments
xx:
The string to send out of the RPM4 COM2 port. It must be less than
40 characters long.
Remarks
The RPM4 COM2 port can be used to communicate to another RS232 device (such as
another RPM4). This allows the user to use a single COM port or IEEE-488 port on the
host computer to communicate with the RPM4 and another RS232 device. A carriage
return and a line feed (
<CR><LF>
) are added to the string.
After this program message is issued, the RPM4 will reply back every string received by
the RPM4 COM2 port that is terminated with a carriage return. Line feeds are discarded.
This will discontinue when the next query is sent to the RPM4.
There is no other reply from this program message. Prior to using this program message,
you must ensure that the RPM4 COM2 port is correctly set up to communicate with the
device on COM2. Refer to the
“COM2=“
program message.
Example
(classic)
Sent:
Reply:
“#VER”
“DH INSTRUMENTS, INC RPM4 us A1000/A0015 Ver2.00 ”
This example assumes that a second RPM4’s COM1 port is connected to the RPM4
COM2 port. This example gets the version of the second RPM4.
See Also
“COM2”
3.5.2
ABORT
Purpose
Cancels any pending data requests by the “PR”, “PRR” “RATE”, “SR” or “CONT”
commands.
Command
Classic
“ABORT”
“ABORT”
Remarks
This program message is usually used to ensure the RPM4 is in an idle condition.
Example
(enhanced)
Cmd sent:
Reply:
“ABORT”
“ABORT” (no reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“ABORT”
“ABORT”
See Also
“PR”, “PRR” “RATE”, “SR” or “CONT”
3.2.1
Содержание RPM4
Страница 10: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VIII N NO OT TE ES S...
Страница 16: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VI N NO OT TE ES S...
Страница 124: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 114 N NO OT TE ES S...
Страница 144: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 134 N NO OT TE ES S...
Страница 148: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 138 N NO OT TE ES S...