3. OPERATION
Page
47
3.3.9.1
[AUTOZ] IN GAUGE AND NEGATIVE GAUGE MODE
PURPOSE
To “rezero” the active Q-RPT in gauge measurement mode.
See Section 3.5.1,
Principle for a full explanation of the AutoZ function.
OPERATION
For the AutoZ function key to run AutoZ, AutoZ must be turned ON for the
active range and measurement mode. AutoZ ON is indicated by <z> to the
left of the Q-RPT designator on the first line of the main run screen. AutoZ
ON and OFF is set using [SPECIAL], <1AutoZ> (see Section 3.5.1). If AutoZ
is OFF for the active Q-RPT, and measurement mode, <AutoZ is off >
is displayed when [AutoZ] is pressed.
Running AutoZ in or parallel mode (see Section 3.5.7.3) autozeros both the
Hi and Lo Q-RPTs simultaneously.
To run AutoZ in gauge measurement mode, set gauge or negative gauge as the
measurement mode (see Section 3.3.3) and press
[AutoZ]
from the main run
screen. If the Q-RPT is a Gxxx or BGxxx type, the display is
<Preparing to
AutoZ>
for five seconds to allow time for the bypass valve between the
TEST(+)
and
TEST(-)
sides of the Q-RPT to assure zero differential pressure and for the
pressure to stabilize (see Section 5.6, Figure 8).
<Running gauge AutoZ>
is
displayed briefly before returning to the main run screen.
Before running AutoZ in gauge mode, ensure that the pressure applied to
the Q-RPT is truly zero gauge (atmospheric pressure). If running AutoZ in
gauge mode results in a zero offset that RPM4 considers unusually large,
<Confirm 0 gauge P!> is displayed when AutoZ is pressed. Check that zero
gauge pressure is applied to the TEST(+) port and there is no differential
pressure across the TEST(+) and TEST(-) ports and press ENTER to
continue or ESCAPE to abort.
Allow the RPM4 to stabilize at atmospheric pressure and ambient
temperature for 1 to 2 minutes before running AutoZ in gauge mode.
Содержание RPM4
Страница 10: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VIII N NO OT TE ES S...
Страница 16: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VI N NO OT TE ES S...
Страница 124: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 114 N NO OT TE ES S...
Страница 144: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 134 N NO OT TE ES S...
Страница 148: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 138 N NO OT TE ES S...