7.1
REMOTE [ENT]
The RPM4 remote ENTER function operates by detecting the open or closed status of the REMOTE
ENTER switch.
The normal status of the ENTER switch is OPEN. When RPM4 detects a CLOSED condition held for 300 ms,
it is interpreted as equivalent to a press of the
[ENT]
key.
Any switch may be used to accomplish the remote ENTER function. The switch should be installed on
the optional remote ENTER cable (
DHI
P/N 103128). Install the switch by connecting its two terminals to
the black and white wires of the remote ENTER cable.
Figure 10.
Remote [ENT] Connector Schematic
1. Black
wire
2. White
wire