RPM4™ OPERATION AND MAINTENANCE MANUAL
Page
54
digit to the right. For example, resolution of 0.001 % on a range of 150 kPa is
150 X 0.001 % = 0.0015 which is rounded down to 0.001 kPa.
Default resolution is 10 ppm of active range span. Resolution is set automatically by
AutoRange (see Section 3.3.4).
OPERATION
To access the resolution function, press
[SETUP]
,
<2res>
. The display is:
Display resltn: Hi
0.0010 %FS < and >
Use the cursor control keys to set the desired level of resolution. Press
[ENT]
to return to the
MAIN RUN screen with the new resolution setting active or
[ESC]
to make no changes.
The resolution setting is range specific. A resolution setting made in one range does NOT
affect other ranges.
The measured pressure resolution is fixed for altitude units at 1 m in meters and 1 ft in feet.
The maximum resolution setting is limited in AutoRange ranges that are less than 10% of the
Q-RPT default full scale.
3.4.3
<3STAB>
PURPOSE
To view and/or adjust the stability test that is the Ready/Not Ready criterion for the active
RPT and range.
See Section 3.2.1 Pressure Ready/Not Ready Indication.
PRINCIPLE
RPM4 continuously monitors the rate of change of pressure measured by the active RPT and
range and compares this rate to the stability limit to make a Ready/Not Ready determination
(see Section 3.2.1 Pressure Ready/Not Ready Indication). The stability function allows the
stability limit to be adjusted by the user to increase or decrease the stability required for a
Ready condition to occur.
The default stability limit is + 0.005% FS of the active range.
The stability limit is separate and different from the Rate DISPLAY function (see Section
3.3.6.2) which allows the current rate of change of pressure to be displayed.
Содержание RPM4
Страница 10: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VIII N NO OT TE ES S...
Страница 16: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VI N NO OT TE ES S...
Страница 124: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 114 N NO OT TE ES S...
Страница 144: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 134 N NO OT TE ES S...
Страница 148: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 138 N NO OT TE ES S...