E110-P
User Manual
FW_E110-P_M_v0301-01_EN
Page 55
Remove old battery from BSM/RSM
1. Disconnect the battery connector cable from
the Supply module.
2. Carefully remove the battery from the holder.
3. Store the old battery in a small plastic bag (e.g.
the bag the new battery came in) or install an
insulation tape over the battery connector to
prevent a short circuit.
4. Dispose of the battery according to the local
environment regulations
Insert new battery in BSM/RSM
1. Press the battery fully into the battery holder
as shown in figure.
2. Carefully click the battery connector cable into
its connection in the Supply module.
7.3.2
REPLACE CLOCK BATTERY (TYPE ZL)
Remove old battery from MEM.
1. Remove the back cover of the MEM by
pushing one clip backwards with a small
screwdriver while pulling the back cover.
Repeat this with the next clip.
2. Carefully lift the back cover and remove
the battery from the battery holder.
3. Dispose of the battery according to local
regulations.
Insert new battery in MEM.
1. Unpack the new battery and ensure there are
no signs of damage or overheating.
2. Press the battery fully into the battery holder
as shown.
3. Carefully click the back cover into its original
position on the MEM.
7.4
REPLACE SILICA GEL SACHET
A silica gel sachet is not supplied with the product but may have been placed during or after
installation to prevent condensation.
Incompatibility risk
Do not use silica gel in environments where hydrogen fluoride, strong acids and/or strong bases are
to be expected
1. Follow the instructions shown in
Section 6.2.3: Opening, assembling and closing the unit [»38]
gain access to the inside of the housing.
2. Remove old silica-gel sachet from housing and replace with a new dry sachet.
3. Dispose of waste in accordance with (inter)national, manufacturer and plant owner standards
and regulations.