Input Interpolation - Input Program
Function
The purpose of these screens in combination provides for linear interpolation of measured analog
input values to be represented as a physical measurement meaning in engineering units.
Navigation
Access these screens by using the CHAN key until reaching the desired port number, and then by
using the PROG key until reaching this screen. Note the first numeric field blinks indicating readiness
for programming.
Select
Field - select value to change by pressing either the QTY (right) or CHAN (left).
Value - change selection by pressing either the ZERO (down) or RATE (up) key.
Program - save new selection by pressing the PROG key.