![Flo Vigaro Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/flo/vigaro/vigaro_instruction-manual_2306435033.webp)
33
Alarms
No.
Display
Error
Continued usage
C9
Internal error
No
C10
Error accessing
memory card
Yes, after
resolving the
error
C15
Heating plate
sensor
indicating a
defect
Yes, with
restrictions
(heating plate
not working)
C16
Internal device
clock faulty
Yes, with
restrictions
(no data
recorded)
C17
Maintenance is
required
Yes
C18
Factory
maintenance is
required
Yes
C19
Motor control
fault
No
C20
Memory card not
supplied by FLO
Medizintechnik
GmbH
No
C22
Memory card
needs to be
reset
Yes, after
resetting the
memory card
using FLOsoft