51
Declarations regarding electromagnetic compatibility
Guidelines and manufacturer's declaration - Electromagnetic immunity
The FLO Vigaro is intended for use in the electromagnetic environment specified below.
The user of the FLO Vigaro should ensure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - Guidelines
Conducted RF
disturbances
as per
IEC 61000-4-6
Radiated RF
disturbances
as per
IEC 61000-4-3
3 V
Effective value
150 kHz
up to 80MHz
3 V/m
80 MHz
up to 2.5 GHz
3 V
Effective value
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the FLO Vigaro incl. cables than the
recommended safety distance calculated from
the equation appropriate to the frequency of
the transmitter.
Recommended safety distance:
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
with
P
is the maximum output power of the
transmitter in watt (W) according to the
transmitter manufacturer and d is the
recommended safety distance in metres (m).
Field strengths from fixed RF transmitters as
determined by an electromagnetic site survey
(a) should be less than the compliance level
(b) in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:
Содержание Vigaro
Страница 1: ...Instruction manual FLO Vigaro Next Generation English Rev 1 0 July 2018 Vigaro from REV 4 15...
Страница 3: ...Instruction manual FLO Vigaro Next Generation English July 2018...
Страница 4: ......
Страница 61: ...61 Notice...
Страница 62: ...62 Notice...
Страница 64: ...Art Nr 40000245...