31
Alarms
8.1.3 Disconnection
Alarm
Disconnection error (C13)
Display
Alarm
No pressure build-up
Delay
Cause of alarm present for at least 10 seconds
Acoustic alarm
Medium priority
whilst cause of alarm present
Visible alarm signal
Yellow LED flashes
Possible cause
- Hose system not connected properly
- Major leakage occurring
Deactivate
The acoustic alarm can be confirmed indefinitely
by pressing the menu key. This deactivation
ceases when the therapy ends or if the cause of
the alarm is no longer present.
Comment
The device does not switch off.
As soon as the fault is corrected, the alarm
disappears from the display, and there is no more
acoustic signal.
8.2
Power failure
If the power supply to the device fails during therapy (e.g. due to an outage or
disconnection of the mains cable), this is indicated by an acoustic alarm
(continuous tone).