Alarms
32
8.3
Low priority alarms
The corresponding fault is displayed for all alarms.
The yellow LED stays lit for all faults.
Device-specific alarms are triggered whenever there is a problem with the
device. With the majority of possible faults, the device can no longer be used
and must be checked immediately by a service engineer. Some alarms can be
switched off using the alarm key (e.g. heating plate faults). This way, and for
the time being, you can continue using the device subject to certain restrictions
(heating plate not working).
Such faults should also be promptly checked by the service team.
No.
Display
Error
Continued usage
C1
Device power
consumption
exceeds
the tol. range
No
C2
Heating plate
power
consumption
exceeds the
tol. range
Yes, with
restrictions
(heating plate
not working)
C5
Internal error
No
C6
Internal error
No
C7
Internal error
No
C8
Pressure
measurement
error
No