User Manual X8400 sc
34
6.5.2 Integrate While Read
The Integration and the Data Readout periods can be thought of as two separate processes.
However, they are linked together by certain timing requirements. This means that the
camera can integrate for a period, starts the data read out for that integration period, and
during that readout starts the integration period for the next frame. This process is called
Integrate While Read (IWR) and can greatly speed up frame rates as seen in
Figure 7
.
The drawback to this process is that it injects a fixed
pattern noise into the data which can be removed by performing a Non Uniformity Correction
(NUC) to the data.
Figure 7: IWR Frame Generation Process
Max Achievable frame rate in IWR Mode – Base Mode
6.5.2.1
Number of points
1280
1152
1024
896
768
640
512
384
256
128
N
u
m
b
e
r
o
f
li
n
e
s
1024
57
63
71
80
92
106
106
106
106
106
896
65
72
80
91
105
121
121
121
121
121
768
76
84
93
106
122
140
140
140
140
140
640
90
100
111
126
145
167
167
167
167
167
512
112
124
138
156
179
206
206
206
206
206
384
148
163
181
205
235
270
270
270
270
270
320
176
193
215
243
278
319
319
319
319
319
256
216
238
265
298
341
391
391
391
391
391
224
245
269
299
337
385
440
440
440
440
440
192
282
310
344
386
441
503
503
503
503
503
160
332
364
404
453
516
587
587
587
587
587
128
403
442
490
548
622
706
706
706
706
706
96
515
563
621
693
784
885
885
885
885
885
64
711
774
850
943
1058
1185
1185
1185
1185
1185
56
786
855
937
1036
1160
1294
1294
1294
1294
1294
48
878
953
1043
1150
1283
1426
1426
1426
1426
1426
40
995
1078
1175
1292
1435
1588
1588
1588
1588
1588
32
1149
1240
1347
1474
1628
1792
1792
1792
1792
1792
24
1358
1459
1577
1716
1882
2055
2055
2055
2055
2055
16
1660
1773
1902
2053
2229
2409
2409
2409
2409
2409
8
2134
2258
2397
2553
2732
2911
2911
2911
2911
2911
2
2717
2841
2976
3125
3290
3449
3449
3449
3449
3449
Table 7: Maximum Frame Rate vs Image Size
(detector Mode: IWR / Base / integration time = 1500µs)
Note
•
In IWR mode, the integration time has much less effect on maximum frame rate.
Integration
Data Read
Frame Sync
Integration Period
Dead Time
Data Width
Dead Time
Integration
Data Read
Frame Sync
Integration Period
Dead Time
Data Width
Dead Time
Integration must end after
the Dead Time following
the previous frame’s Data
Read
Integration can’t start
before previous frame’s
Data Read
Содержание X8400sc
Страница 1: ...User Manual X8400 sc 1 X8400sc User Manual Document Number DC085U Version C Issue Date 08 03 2013...
Страница 42: ...User Manual X8400 sc 42 4 Connect to http 192 168 64 1...
Страница 48: ...User Manual X8400 sc 48 7 3 Mechanical General Below are the mechanical interfaces...
Страница 49: ...User Manual X8400 sc 49...