Manual MCS-64 Page 43
TL
Trigger
Level
[ 2500sub01 ]
Set the trigger level for rising edge start of measurement. Permitted values are in the
range 0…262143.
Master (PC / PLC) sends
Device responds
Result
TL T+99999
Request: TL=99999
TL_1000 OK
Setting: TL=1000
With regard to the trigger commands SD and MT, a check weighing will automatically start
when the weight overshoots by e.g. 1.000d (increments), e.g. 100,0 g.
Factory default: 199999
TE Trigger
Edge
[ 2500sub02 ]
Issuing the TE command selects rising or falling edge trigger. Parameter = 0 select falling
edge and parameter = 1 select rising edge.
Master (PC / PLC) sends
Device responds
Result
TE E:001
Request: TE=1
TE_0 OK
Setting: TE=0
Factory default: 0 [falling edge]
Note:
This command cannot be used together with the TL command.
TR
Trigger
RPDO2 [ 00 80 ]
[This command will start the measuring cycle in the same way as the hardware trigger
input. The result is sent as a TPDO2 with source=LDM#.]
Master (PC / PLC) sends
Device responds
Result
TR OK
Trigger started
Note:
This function can be used as a soft trigger in a check weigher application (firmware
88.183) or to start a filling process (firmware 88.184)
[ Special returned values are:
- 99999 = trigged measurement in progress
- 88888 = Re-Trigger in progress
- 99996 = Tried to change Tare or Zero before end of measurement ]