Page Manual MSC-64
46
DT
Delta Time
[ 2500sub0B ]
Set the Delta Time in milliseconds [ms].
During MT and TT timeframes "sub-averages" will be
calculated by the system over the time DT. If a sub-average is outside the re-trigger window,
the re-trigger function is automatic started.
Master (PC / PLC) sends
Device responds
Result
DT T+00000
Request: DT=00000
DT_50 OK
Setting: TT=50ms
Factory default: 00000
TW
Tare Window
[ 2500sub06 ]
Set the Tare Window in in counts (digits) without decimal point. Tare window (TW) allows an
automatic Tare update. If TW = 0 this function is not active. If TW = 100, this means a new
tare value will be taken when the net average weight of an empty scale is within 100 counts
or division of zero. The new average tare value is calculated over the average tare time
defined by TI. If the tare average is outside tare window, the tare will not be updated.
Master (PC / PLC) sends
Device responds
Result
TW T+00000
Request: TW=00000
TW_100 OK
Setting: TW=100d
Factory default: 00000
TI
Tare Time
[ 2500sub07 ]
Set the Tare Time in milliseconds [ms].
During the Tare Time a “tare-average” will be
calculated by the system.
Master (PC / PLC) sends
Device responds
Result
TI T+00000
Request: TI=00000
TI_200 OK
Setting: TI=200ms
Factory default: 00000
HT
Hold Time
[ 2500sub05 ]
Set the Hold Time in milliseconds [ms].
During the Hold Time the weight value must be over
setpoint that a digital output can be switched. This means that a short signal peak will not
lead to the switching of a digital output.
Master (PC / PLC) sends
Device responds
Result
HT T+00000
Request: HT=00000
HT_200 OK
Setting: HT=200ms
Factory default: 00000