9 Damping mats
72
UMFLUXUS_G8V4-6-1EN, 2018-10-10
Example:
measurement in diagonal arrangement
2 transducer mounting fixtures Variofix C
G transducer
width of the damping mat: 225 mm
outer pipe diameter: 1200 mm
width of Variofix C: 560 mm
reflection points: 2
Calculation of the installation lengths (see Tab. 9.2):
• transducer damping mat: l
S1
= 600 mm
• pipe damping mat: l
R
= 600 mm (2 x)
Total installation length is 1800 mm.
Transducer damping mats are installed on the pipe lengthwise. The pipe damping mats can be installed
on the pipe lengthwise or crosswise. In the example, they are installed crosswise.
Number of transducer damping mats
The transducer damping mats are installed along the pipe circumference with a width of 2 x 1400 mm.
The number of transducer damping mats is 2 x 1400 mm / 225 mm = 2 x 6.2.
2 x 6 damping mats (600 mm x 225 mm) + 2 x 1 damping mat (600 mm x 50 mm)
Number of pipe damping mats
The pipe damping mats are installed along the entire pipe circumference.
pipe circumference: 2
π
r = 3770 mm
The number of pipe damping mats to be installed is 600 mm / 225 mm (2x). The value is rounded up.
2 x 3 pipe damping mats (3770 mm x 225 mm)
The damping mats can be cut into smaller pieces to make installation easier.
l
R
reflection point
pipe
damping
mats
transducer
damping
mats
pipe
damping
mats
outer pipe diameter > 900 mm
l
R
l
S1
reflection point