A Menu Structure
140
UMFLUXUS_G8V4-6-1EN, 2018-10-10
Current loop
activation of a current output
This display will only be indicated if the current output has been installed in
Special Funct.\SYSTEM settings\Proc. outputs
.
selection whether the sign of the measured values is to be considered for the
output
This display will only be indicated if
Current Loop
is activated.
input of the lowest/highest measured value to be expected for the current out-
put
The values are assigned to the lower/upper limit of the output range.
These displays will only be indicated if
Current Loop
is activated.
input of the error value delay, i.e. the time interval after which the error value
is transmitted to the output in case no valid measured values are available
This display will only be indicated if
Special Funct.\SYSTEM set-
tings\Dialogs/Menus\Error-val. delay
=
EDIT
is selected.
Pulse output
activation of a Pulse Output
This display will only be indicated if a pulse output has been installed in
SYS-
TEM settings
\
Proc. outputs
\\.
input of the pulse value (totalizer value at which a pulse will be emitted)
This display will only be indicated if
Pulse Output
is activated.
input of the pulse width
range: 1...1000 ms
This display will only be indicated if
Pulse Output
is activated.
Alarm output
Activation of an Alarm Output
This display will only be indicated if an alarm output has been installed in
Special Funct.\SYSTEM settings\Proc. outputs
.
Selection of the switching condition (
func
), the holding behavior (
typ
) and
the switching function (
mode
) of the alarm output.
This display will only be indicated if
Alarm Output
is activated.
selection of the physical quantity to be monitored
This display will only be indicated for R1 if
Alarm Output
is activated.
input of the upper limit of the physical quantity to be monitored
This display will only be indicated if
Alarm Output
is activated and
MAX
is
selected as the switching condition.
input of the lower limit of the physical quantity to be monitored
This display will only be indicated if
Alarm Output
is activated and
MIN
is
selected as the switching condition.
INIT-resis-
tant
Current Loop
I1: no
>YES<
Meas.Values
>ABSOLUT<
sign
Zero-Scale Val.
0.00
m3/h
Full-Scale Val.
300.00
m3/h
Error-val. delay
10
s
Pulse Output
B1: no
>YES<
Pulse Value
0.01
m3
Pulse Width
100
ms
Alarm Output
no
>YES<
R1=FUNC<typ mode
Function:
MAX
R1 Input:
↕
Volume flow
R1 Input:
↕
act. Volume flow
High Limit:
-10.00
m3/h
Low Limit:
-10.00
m3/h