9 Damping mats
70
UMFLUXUS_G8V4-6-1EN, 2018-10-10
Example:
measurement in reflection arrangement
2 transducer mounting fixtures Variofix L
M transducer
width of the damping mat: 50 mm
outer pipe diameter: 100 mm
length of Variofix L: 310 mm
reflection point: 1
calculation of the installation length (see Tab. 9.1):
• transducer damping mat: l
S2
= 660 mm
• pipe damping mat: l
R
= 350 mm
Total installation length is 1010 mm.
Transducer damping mats are installed on the pipe lengthwise. The pipe damping mats can be installed
on the pipe lengthwise or crosswise. In the example, they are installed lengthwise.
Number of damping mats
The damping mats are installed on the entire pipe circumference.
pipe circumference: 2
π
r = 315 mm
The number of installed damping mats is 315 mm / 50 mm = 6.3.
6 damping mats (1010 mm x 50 mm) + 1 damping mat (1010 mm x 15 mm)
The damping mats can be cut into smaller pieces to make installation easier.
l
R
l
S2
reflection point
outer pipe diameter < 900 mm
transducer
damping
mats
pipe
damping
mats