www.eu.fishersci.com/fisherbrand
8
1.4 Operation
Accu20 water systems produce ultrapure water and RO water directly
from tap water. Tap water is pretreated through a Pretreatment Cartridge
(AccuDuo P Pack) to reduce free chlorine, larger particles and water
hardness. Then most particles, ions and organic compounds are removed
through the RO membrane. RO water flows through an ultraviolet UV light
chamber (optional) to kill bacteria and destroy trace organic pollutants in
water. Then water flows through a polishing cartridge U pack to remove
the last traces of ions, finally, a 0.2 µm filters to the outlet.
After powering-up, press START button to initiate the system. After a
180-second automatic RO flush, the system goes into READY mode.
Press the RO button to start a 60-second cleaning cycle and dispense
RO water. Press the RO button again to stop the dispensing. Press the
UP button to start a 60-second cleaning cycle and dispense ultrapure
water. Press the UP button again to stop the dispensing.
Содержание 15489689
Страница 1: ...www eu fishersci com fisherbrand Accu20 Ultrapure Water System USER MANUAL...
Страница 14: ...www eu fishersci com fisherbrand 14 2 3 5 Accu20 Water Flow Diagram...
Страница 39: ......