4.3 Replacing the RO Membrane
Warning!
RO membrane contains NaHSO
3
as preservative. It may cause
irritation to the mucus membrane. Be careful not to get into the eyes!
If solution inadvertently gets into the eyes, immediately flush eyes
with a large amount of water. If you still feel uncomfortable, please
seek medical attention!
Caution!
RO reverse osmosis membrane should be replaced by a fully trained
professional technician.
1) Switch off the system power, unplug the power cord.
2) Remove the system left side panel. RO membrane housing is
indicated in the picture above.
3) Remove the RO membrane housing from its support.
RO DRAIN
RO OUT
RO IN
RO Housing
Cap
RO Housing
Support
RO Housing