www.eu.fishersci.com/fisherbrand
28
4.1 Replacing the AccuDuo P Pack, U Pack Cartridges and the
Final Filter
Attention!
Do not mix up cartridge positions!
These two cartridges work in different ways. These two packs must be
installed in its proper position as indicated in the picture below. Never
reverse the positions of these packs.
1) Press the START button to put system into Standby. Power off the
system.
2) Screw off the final filter and remove the system right side panel
3) Remove the used cartridges by lifting up the lock chip, and then pull
the cartridges out.
4) Remove blue caps on the new cartridges.
5) Install P pack and U pack as the procedures described in 2.4.1.
6) Degassing the system and check system leakage
a) Power up the system, then press the START button.
b) Press the UP button to degas the system and check for system
leakage. If there is a leak, turn off the system and reinstall the
cartridge.
c) Press the UP button to put system into Ready.
P Pack
U Pack