Page 15
Isotemp Economy Incubators
Trouble-Shooting Table
This table is intended to assist in the resolving incubator problems by
relating symptoms to their likely causes. The service discussed below is
beyond the scope of most users and should be performed by qualified and
trained personnel.
Symptom Probable
Cause
Action
No power
Unit not plugged in or turned on.
Plug in or turn on.
Defective circuit breaker.
Replace circuit breaker.
Incubator temperature
Defective control thermocouple
Replace control thermocouple
erratically high
Failure to heat
Set temperature less than
Refer to
Operation Section
actual temperature
Defective control thermocouple
Replace control thermocouple
Poor
heater
connections
Tighten connections at terminal strip
Defective heater element
Check heater resistance on schematic
at back of manual. Replace heater
unless approximately same as
schematic.
Defective controller
Replace controller
Alarm LED stays on and Set temperature has been
Wait for actual temperature to cool to
control is higher than set changed to a value less than
the set temperature
temperature
the high alarm limit
Defective control thermocouple
Replace control thermocouple
Defective controller
Replace controller