Page 5
Isotemp Economy Incubators
Safety
Precautions
Before Operating the oven, always
observe the
following Safety Precautions:
This unit is not explosion proof,
do not use in the presence of
flammable or combustible materials; fire or explosion may result. Unit
contains components, which may ignite such materials.
Fumes and spillage from acidic solutions cause corrosion of the
stainless chamber. Care should be taken to maintain a neutral pH at all
times.
•
Wear
insulated
gloves.
•
Use tongs.
•
Never stand in front of an open oven.
•
Use safety goggles.
Operation
The 500 Series incubators maintain a set temperature until that set
temperature is changed.
To achieve a set temperature, perform the following:
1. Rotate the temperature dial full counter-clockwise.
2.
Place the power switch in the ON position. The Power indicator will
come on.
3. Rotate the temperature dial to the desired set temperature. The Heat
indicator will then come on.
4.
When the actual temperature approaches the set temperature, the
Heat indicator will cycle on and off, maintaining the set temperature.
NOTE
: 1. After controlling at a given temperature, if the
temperature dial is rotated CCW, the alarm indicator will
come on. This is normal operation.
2. Since this is an analog control with 5 degree
increments. The user may choose to add a thermometer
and make minor adjustments to reach a more precise
temperature setting.
Содержание Isotemp 500 Series
Страница 2: ......
Страница 7: ......
Страница 23: ...Page 19 Isotemp Economy Incubators Electrical Schematic ...