Isotemp Economy Incubators
Page 8
7. Carefully remove the temperature dial knob without rotating it.
8. Use a thin blade screwdriver to calibrate Low Adjust. Check
thermometer reading. If reading is less than 39
°
C, rotate Low
Adjust slightly to turn heater off at lower temperature. After each
adjustment allow sufficient time for incubator temperature to re-
stabilize. Repeat until thermometer reading is 40
°
C
±
1
°
C.
9. Replace temperature dial knob with pointer at 40
°
C, then tighten set
screw, being careful not to rotate the dial shaft.
10. Set temperature dial to 70
°
C
11. After the Heat indicator begins cycling on/off, allow an additional 30
minutes for incubator to reach a final, stable temperature. If
thermometer reads 70
°
C +/- 1
°
C, calibration is complete.
Otherwise continue with the following steps.
12. Carefully remove the temperature dial knob without rotating it.
13. Use a thin blade screwdriver to calibrate High Adjust. Check
thermometer reading. If reading is less than 69
°
C, rotate High
Adjust slightly to turn heater off at lower temperature. After each
adjustment, allow sufficient time for incubator temperature to re-
stabilize. Repeat until thermometer reading is 70
°
C +/- 1
°
C.
14. Replace temperature dial knob with pointer at 70
°
C, then tighten
set screw, again being careful not to rotate the dial shaft.
15. Set the temperature to 60
°
C. The Alarm indicator should come on
if calibration is correct.
Содержание Isotemp 500 Series
Страница 2: ......
Страница 7: ......
Страница 23: ...Page 19 Isotemp Economy Incubators Electrical Schematic ...