Troubleshooting
48
Error
Symptom
Possible causes and corrective measures
E-22/E-23
Rotor stops without
deceleration to standstill.
Instrument cannot be
operated.
Error in speed entry
Switch the instrument off and on again.
If the error persists, contact service representative.
E-24
Instrument cannot be
operated.
Wrong status information from the lid-latch.
1. Turn the centrifuge off and on.
2. After re-switching on, the display shows "
Lid FAiL
"
3. If the centrifuge lid has been already opened, the display
shows "
Close Lid
". Thereupon close the lid.
4. The centrifuge tries to open the lid to switch for starting
the normal operation mode.
If the error persists, contact service representative.
E-29
Motor does not start.
Motor or rotor blocked.
1. Switch instrument off and on again using the mains
switch.
2. Open the centrifuge lid.
3. Check whether the rotor can turn freely.
If you cannot clear the malfunction, contact a service
representative.
Содержание AccuSpin Micro 17
Страница 1: ...Instructions for use AccuSpin Micro 17 17R ...
Страница 10: ...For your safety 6 for your notes ...
Страница 46: ...Maintenance and care 42 ...
Страница 56: ...Troubleshooting 52 for your notes ...
Страница 62: ...Technical data 58 for your notes ...
Страница 63: ...Appendix 59 Appendix ...
Страница 69: ...Appendix 65 For your notes ...
Страница 70: ...Appendix 66 Autoclaving protocol Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10 ...
Страница 71: ...Appendix 67 Autoclaving protocol Date Remark Operator Signature 11 12 13 14 15 16 17 18 19 20 ...
Страница 72: ...Appendix 68 For your notes ...
Страница 78: ...74 for your notes ...