Troubleshooting
47
Error
Symptom
Possible causes and corrective measures
„bAL“
The rotor stops with decel-
eration.
Imbalance switch releases.
1. Open the instrument by pressing the “lid open”
in case.
2. Check whether the rotor is properly loaded.
3. Check whether a broken tube or a damage to the rotor
release the imbalance switch.
If you cannot locate and solve the problem yourself, contact a
service representative.
E-01
|
E-13
Rotor stops without
deceleration to standstill.
Instrument cannot be
operated.
Internal program error.
Switch the instrument off and on again.
If the error persists, contact service representative.
E-14
Rotor stops with
deceleration to standstill.
Instrument cannot be
operated.
Overtemperature in the centrifuge chamber.
Switch the centrifuge off and turn it on again after approx. one
minute.
If the error persists, contact service representative.
E-15
|
E-16
Rotor stops with
deceleration to standstill.
Instrument cannot be
operated.
Temperature measurement error.
Switch the instrument off and on again.
If the error persists, contact service representative.
Содержание AccuSpin Micro 17
Страница 1: ...Instructions for use AccuSpin Micro 17 17R ...
Страница 10: ...For your safety 6 for your notes ...
Страница 46: ...Maintenance and care 42 ...
Страница 56: ...Troubleshooting 52 for your notes ...
Страница 62: ...Technical data 58 for your notes ...
Страница 63: ...Appendix 59 Appendix ...
Страница 69: ...Appendix 65 For your notes ...
Страница 70: ...Appendix 66 Autoclaving protocol Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10 ...
Страница 71: ...Appendix 67 Autoclaving protocol Date Remark Operator Signature 11 12 13 14 15 16 17 18 19 20 ...
Страница 72: ...Appendix 68 For your notes ...
Страница 78: ...74 for your notes ...