Maintenance and care
41
Decontamination
For general radioactive decontamination, use a solu-
tion of equal parts of 70% ethanol, 10% SDS and wa-
ter. Follow this with ethanol rinses, then de-ionized
water rinses, and dry with a soft absorbent cloth. Dis-
pose of all washing solutions in appropriate radioactive
waste containers!
Autoclaving
Check whether autoclaving is permitted!
You may autoclave the rotor, rotor lid and the adapters
at 121 °C.
Maximum permissible autoclaving cycle:
20 minutes at 121 °C.
For safety reasons, the plastic
rotors and rotor lids must only
be subjected to a maximum of 20
autoclaving cycles!
The rotor must be cleaned and rinsed with distilled
water before being autoclaved. Remove the rotor lid,
the centrifuge tubes and the adapters. Place plastic
rotors on an even surface to avoid deformation.
Chemical additives to the steam are not
permitted.
Never exceed the maximum permis-
sible values for autoclaving tem-
perature and autoclaving time.
Should the rotor show signs of wear
or corrosion, you must stop using
it!
Содержание AccuSpin Micro 17
Страница 1: ...Instructions for use AccuSpin Micro 17 17R ...
Страница 10: ...For your safety 6 for your notes ...
Страница 46: ...Maintenance and care 42 ...
Страница 56: ...Troubleshooting 52 for your notes ...
Страница 62: ...Technical data 58 for your notes ...
Страница 63: ...Appendix 59 Appendix ...
Страница 69: ...Appendix 65 For your notes ...
Страница 70: ...Appendix 66 Autoclaving protocol Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10 ...
Страница 71: ...Appendix 67 Autoclaving protocol Date Remark Operator Signature 11 12 13 14 15 16 17 18 19 20 ...
Страница 72: ...Appendix 68 For your notes ...
Страница 78: ...74 for your notes ...