Troubleshooting
46
Error
Symptom
Possible causes and corrective measures
–
Exceptionally running
noise.
Imbalance.
1. Stop the centrifuge by pressing the "stop" key, in case of
emergency, unplug mains power cord.
2. Wait until the centrifuge comes to a complete stop.
3. Check whether the rotor is properly loaded.
4. Check whether a broken tube, damage to the rotor or
motor is responsible for the run noise.
If you cannot locate and solve the problem yourself, contact a
service representative
Display "oP" appears
although lid is closed
.
Will not start.
Centrifuge lid not properly closed
- Open the lid and repeat locking procedure.
If you cannot locate and solve the problem yourself, contact a
service representative.
“Lid” Rotor
stops
with
deceleration to standstill.
Centrifuge lid was opened manually during the run.
- Close centrifuge lid immediately
The instrument phases out with deceleration.
For further centrifugation, you have to switch the instrument
off and switch it on again.
Содержание AccuSpin Micro 17
Страница 1: ...Instructions for use AccuSpin Micro 17 17R ...
Страница 10: ...For your safety 6 for your notes ...
Страница 46: ...Maintenance and care 42 ...
Страница 56: ...Troubleshooting 52 for your notes ...
Страница 62: ...Technical data 58 for your notes ...
Страница 63: ...Appendix 59 Appendix ...
Страница 69: ...Appendix 65 For your notes ...
Страница 70: ...Appendix 66 Autoclaving protocol Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10 ...
Страница 71: ...Appendix 67 Autoclaving protocol Date Remark Operator Signature 11 12 13 14 15 16 17 18 19 20 ...
Страница 72: ...Appendix 68 For your notes ...
Страница 78: ...74 for your notes ...