![Fisher Scientific AccuSpin Micro 17 Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/fisher-scientific/accuspin-micro-17/accuspin-micro-17_instructions-for-use-manual_2296621019.webp)
Accessories
15
Table 3: Performance characteristics for accuSpin Micro 17R
rotor designation
order no.
PCR-rotor 4 x 8
75003440
PCR-rotor 8 x 8
75003489
Haematocrit-rotor
75003473
places / volume
4 x PCR-Strip
8 x PCR-Strip
24 x Blood capillary tubes
75 mm
maximum permissible load [ g ]
4 x 4
8 x 4
24 x 0,2
minimum speed n
min
[ min
-1
]
300
300
maximum speed n
max
[ min
-1
]
13 300
13 300
13 300
maximum RCF value at n
max
13 100
13 800
16 800
acceleration / deceleration time [ s ]
9 / 12
6 / 8
9 / 11
radius maximum / minimum [ cm ]
6,6 / 4,7
7,0 / 4,4
2,0 / 8,5
angle [ ° ]
45
60
90
minimum temperature at n
max
[ °C ]
relative to room temperature 23°C
< 0
< 0
< 0
aerosol-tight *
yes
no
no
permissible temperature range
autoclavable (number of cycles)
-4 °C to +40 °C
121°C, (20 cycles)
-4 °C to +40 °C
121°C, (20 cycles)
-4 °C to +40 °C
134°C
* Tested and approved by HPA, Porton-Down, UK – Please see the references in the chapter "Aerosoltight application"!
Содержание AccuSpin Micro 17
Страница 1: ...Instructions for use AccuSpin Micro 17 17R ...
Страница 10: ...For your safety 6 for your notes ...
Страница 46: ...Maintenance and care 42 ...
Страница 56: ...Troubleshooting 52 for your notes ...
Страница 62: ...Technical data 58 for your notes ...
Страница 63: ...Appendix 59 Appendix ...
Страница 69: ...Appendix 65 For your notes ...
Страница 70: ...Appendix 66 Autoclaving protocol Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10 ...
Страница 71: ...Appendix 67 Autoclaving protocol Date Remark Operator Signature 11 12 13 14 15 16 17 18 19 20 ...
Страница 72: ...Appendix 68 For your notes ...
Страница 78: ...74 for your notes ...