24
Report: A diagnostic report can be output to either the serial port or to the serial printer (sold sepa-
rately), which summarizes the information contained within the Diagnostics menu. See below for an
example report.
3.6 MANUAL
CALIBRATION
The information contained in this section refers to the Manual Calibration section of the software. To
complete a full manual calibration refer to Section 7.
Enter Password: Prior to performing any calibration on the instrument it is necessary to enter the
password (417). This is to prevent unauthorized access which may result in a reduction in perfor-
mance of the instrument.
Refer to Section 6.2, Manual Calibration - Ambient Temperature
Temp.: Calibration of the Temperature sensor. This is normally factory set and should not require
any adjustment
Version No. V1.00H
Options
Serial Port
Data Log
Pressure Transducer
Ranges
CO
2000
O
2
2.5
NO (FT 1300-1 Only)
4000
SO
2
(FT1200-2A Only)
2000
Gains
CO
-2500
NO(FT 1300-1 Only)
-3750
SO
2
(FT1200-2A Only)
593.54
Draft
4.5
Zeros:
CO
-0.010V
O
2
-0.004V
NO (FT 1300-1 Only)
0.000V
SO
2
(FT1200-2A Only)
0.000V
ppm. days:
CO
0
NO (FT 1300-1 Only)
0
SO
2
(FT1200-2A Only)
0
EXIT
F1
F2
F3
F4
HELP
REPORT
GASES
SYSTEM REPORT
SYSTEM OUTPUT DEVICE: SERIAL PORT
EXIT
F1
F2
F3
F4
HELP
GASES
TEMP: CO
DRAFT ZERO NO (FT1300-1 ONLY)
DRAFT SPAN
MANUAL CAL