47
dBASE III TABLE
The following table shows the field names within type dBASE III database table generated by the
CAPTURE program along with the field types and field widths.
10.6
FAULT DIAGNOSIS
If the CAPTURE software generates a fault message indicating that no data was received then the
following table is a list of the possible cause of failure and the recommended action to be taken.
Fault Description
The CAPTURE program fails to receive the logged data.
Field name
Data Type
Field Width
Date
Date
8
Hours
Numeric
2,0
Minutes
Numeric
2,0
Seconds
Numeric
2,0
T_AMB
Numeric
6,0
T_GAS
Numeric
6,0
CO
Numeric
6,0
O
2
Numeric
6,2
NO
Numeric
6,0 (FT1300-1 Only)
CO
2
Numeric
6,2
NOx
Numeric
6,0 (FT1300-1 Only)
EFF
Numeric
6,1
XS_AIR
Numeric
6,1
WATER
Numeric
6,1
NORM_VAL
Numeric
6,1
ANALYSIS
Character
3
O
2
_NORM
Character
3
UNIT NO.
Numeric
2,0
FUEL
Numeric
1,0
CAUSE
ACTION
The serial port is already in use
Wait until the port is free
The serial port on the Series 1000 is not setup correctly Check under Setup Serial Port
The serial cable is wired incorrectly
Check
The wrong PC port is being used.
Use COMI (or COM2)
There is no logged data on the Firetron 1000.
Check under DATA LOG on the main menu
You are using the continuous CAPTURE mode an the
Firetron 1000 is not set up for transmission to the
RS232 Port
Check under DATA LOG or Continual Monitoring set up