FST-DCP-1500-05 HIGH VOLTAGE POWER SUPPLY
2283
USER MANUAL
22/01/2010 12:52 PM
AT
UserManual FST-DCP-1500-05 HighVoltagePowerSupply 100122EN
ver1.doc
24/27
------------------------------------------------------------------
Finero Oy
–
The Quality Control Company
Tel. +358-5-7522 200
Haltonintie 5, FIN-47400 KAUSALA, FINLAND
Fax +358-5-3638 326
www.qacontrol.com
OUT1:ON
External control output 1 on
OUT2:ON
External control output 2 on
OUT1:OFF
External control output 1 off
OUT2:OFF
External control output 2 off
IDN ” ”
Tester identify code (free setting), maximum 20 characters (not ”
-
marks). Normal there are tester type and serial number.
Example:
IDN ”FST
-
DCP sno 12345”
RES?
Give test results: leakage current, test voltage, status byte 1, test
time. Return answer four numbers : aa bb cc dd
aa
Current (0…1100), 10999=range high limit,
-1= can
not measure
bb
Test voltage (0-2000)
cc
Status byte 1 (see STAT?)
dd
Test time from measure starting (0..9999=0..999min)
Example:
DCP;VOLT 500
Commands to power, test voltage 500V,
CURR:MAX 100
current high limit 100mA, test
ON
voltage on and measurements starts.
RES?
Test results inquiry
OFF
Test stop
CURR?
Test voltage, return answer in real voltage in test. (0…1100)