FST-DCP-1500-05 HIGH VOLTAGE POWER SUPPLY
2283
USER MANUAL
22/01/2010 12:52 PM
AT
UserManual FST-DCP-1500-05 HighVoltagePowerSupply 100122EN
ver1.doc
17/27
------------------------------------------------------------------
Finero Oy
–
The Quality Control Company
Tel. +358-5-7522 200
Haltonintie 5, FIN-47400 KAUSALA, FINLAND
Fax +358-5-3638 326
www.qacontrol.com
8.5 Tester functions but high voltage is missing or voltage output is wrong
There may be overload and then the temperature switches inside of the transformer are
opened. Temperature switches are automatic resetting, only waiting some time. The reason
for the transformer temperature state must be investigated.
8.6 Tester functions but “External control”
-connections not working
Check fuses F1 in external control board. Board is inside tester in back panel.
If the above remedies do not help, contact the Tester service.