FST-DCP-1500-05 HIGH VOLTAGE POWER SUPPLY
2283
USER MANUAL
22/01/2010 12:52 PM
AT
UserManual FST-DCP-1500-05 HighVoltagePowerSupply 100122EN
ver1.doc
23/27
------------------------------------------------------------------
Finero Oy
–
The Quality Control Company
Tel. +358-5-7522 200
Haltonintie 5, FIN-47400 KAUSALA, FINLAND
Fax +358-5-3638 326
www.qacontrol.com
Example:
RAMP 1
Voltage increase slow and decrease fast
RLEN 15
Voltage increase time are 15s
RLEN 0
Maximum speed
ON
Test voltage on and measurement starting. (same like Start-
button). Remove same time ALAR:ON command effect. Reset
some status bits (see STAT?).
Example:
DCP;VOLT 500
Commands to power, test voltage 500V,
CURR:MAX 100
current high limit 100 mA, test
ON
voltage on and measurements starts.
OFF
Test voltage off. (same like Stop-
button). If test voltage are “ON”
before command, then voltage goes off (test ok
–
display). If test
are off before command, tester display setting values (ready
–
display). Reset some status bits (see STAT?).
Example:
HVT;VOLT 500
Commands to power, test voltage 500V,
CURR:MAX 100
current high limit 100mA, test
ON
voltage on and measurements starts.
Test time loop
OFF
Test stop
ALARm:ON
Stops testing and set Fail indicator on and buzzer ringing about 1
seconds. Used for software fail- alarm.
Example:
ALAR:ON
ALARm:OFF
Tester Fail indicator off
Example:
ALAR:OFF