Tester commands, specification and examples:
DCP
All next commands are use to FST-DCP tester (high voltage
power).
Example:
DCP;VOLT 1000;ON
Command to power and test voltage
1000V and test voltage output ON
VOLTage mmm
Setting test voltage (0V…2000V).
mmm= test voltage
Examples:
VOLT 1000
Test voltage 1000V
VOLT 240
Test voltage 240V
CURRent:MAXimum
Setting current high limit value (10...990). (=10..990mA)
mmm=current high limit
If measurement value exceed high limit, voltage increasing stops
and “current limit” goes display and timer counts
20s
0s. If cur-
rent goes lower, voltage are increased to set value. If alarm are
20s, voltage goes off and tester makes alarm and status 1 bit 1
(test fail) and status 2 bit 0 (20s timeout) and bit 3 (current limit)
settings up.
Examples:
CURR:MAX 100
Current high limit 100mA
CURR:MAX 990
Current high limit 990mA
RAMPtype m
Test voltage increase/decrease style 1 or 2. 1=voltage connected
m=style
slow (increase are controlled) and unconnected to direct to 0.
2=both are controlled.
RLENgth mm
T
est voltage increase/decrease time (0)4…30 seconds.
mm=time