4 INSTALLATION 

 

4.1 General 

 

The Power Supply is designed for standard 19" rack or case. Height is 3U (133 mm). Depth is 

about 370 mm and connectors need about 150mm space from rear. The handles mounted on 
the front panel are elevated about 40 mm from the panel. The Power Supply is not allowed to 

carry from the handles but by having a stable hold in the frame. 

 

The  Power  Supply  can be  used  in  its  own  case  only  in  dry  and  clean  environment  but  care 
must  be  taken  nothing  goes  into  the  case  through  ventilation  openings.  The  ventilation, 

however, should not be prevented. 

 

4.2 Mounting into 19” module –

rack or case 

 

When installed into a rack or a similar case it is mandatory to use mounting rails due to the 

weight  of  the  device.  The  holes  in  the  front  panel  are  intended  for  locking  the  device  in  its 

place  not  for  backing  the  whole  device.  If  there  are  legs  below  of  tester,  they must  take  off 

before installation to 19” rack. Care must be taken of the air circ

ulation in the rack in order to 

avoid temperature rise above +35°C. 

 

Here are some examples of how build the support for rack installation. 
 

 

 

 

 
 

Содержание DCP-1500-05

Страница 1: ...ower Supply DCP 1500 05 Generalvertretung Deutschland TESTEC Elektronik GmbH Westerbachstr 58 60489 Frankfurt am Main Telefon 49 0 69 943335 0 Telefax 49 0 69 943335 55 e mail info testec de Internet...

Страница 2: ...voltage 8 4 INSTALLATION 9 4 1 General 9 4 2 Mounting into 19 module rack or case 9 4 3 High voltage connectors 10 4 4 External Safety Circuit Emergency Switches 10 4 5 Mains supply 10 5 MANUAL OPERAT...

Страница 3: ...but External control connections not working 17 9 LIMITED WARRANTY 18 ANNEX A FST DCP 1500 05 HIGH VOLTAGE POWER SUPPLY COMMAND LIST Any non authorized modification repair tampering or physical damage...

Страница 4: ...nero test and measurement equipment Finero is obliged to provide you with the best on the market available test and measurement equipment The 30 years of Finero experience in most demanding production...

Страница 5: ...ation is subject to changes without prior notice DELIVERY SCOPE LIST Although Finero packages the equipment with utmost care you should check the package regarding any delivery damages If the equipmen...

Страница 6: ...s 4A slow 5x20mm glass inside of power connector Environment temperature 5 C 35 C Environment conditions Dry and no dusty Output voltage 0V 2000V dc voltage not filtered only rectifier Meter 0 2000V a...

Страница 7: ...omatic test systems by a computer via RS 232 or IEE488 or USB connection Also logic or remote control of basic functions is possible 2 3 Construction The Power Supply is built in a 19 frame having hei...

Страница 8: ...00 05 HighVoltagePowerSupply 100122EN ver1 doc 7 27 Finero Oy The Quality Control Company Tel 358 5 7522 200 Haltonintie 5 FIN 47400 KAUSALA FINLAND Fax 358 5 3638 326 www qacontrol com sales qacontro...

Страница 9: ...d before starting any tests As safety precaution the operator should be aware of emergency first aid actions 3 2 High voltage The high voltage connectors in rear panel have a dangerous high voltage du...

Страница 10: ...to carry from the handles but by having a stable hold in the frame The Power Supply can be used in its own case only in dry and clean environment but care must be taken nothing goes into the case thr...

Страница 11: ...when circuits are open The safety circuit connector is a part of the external connec tor The wiring diagram is shown in tester wiring diagram When the safety circuit is open the Fail indicator is flas...

Страница 12: ...switch Dating information is shows few seconds in display panel and then comes tester set values Ready display Power Supply program versions are possible seen if pressed Enter button when calibration...

Страница 13: ...Remote EXIT Menu commands U set Voltage setting 0 2000V I max Leakage current high limit value 10mA 990mA Ramp Voltage increase method up ramp or 2 ramp and ramp time 0 4 30s Save Setting values store...

Страница 14: ...ror displays Power Supply make alarms if measurements value are not inside limits Errors CURRENT LIMIT Current is higher than limit I max Tester stops volt age increasing and starts 20s timer Output v...

Страница 15: ...a specific application program Finero Safetest The command lists for computer operation mode are appendix of this manual The control board RS 232 or IEE488 or USB are accessory and its must be ordere...

Страница 16: ...e In addition the wiring of the operator connector at the rear is to be done The connector is a 25 pin D type female connec tor In wiring diagram you ll find a diagram and wiring of the equipment whic...

Страница 17: ...eplace with the similar one if neces sary 8 2 No indicator is lit Check the power cord connectors are properly plugged at both ends Check the voltage in the mains socket Check the main fuses F1 and F2...

Страница 18: ...5 Tester functions but high voltage is missing or voltage output is wrong There may be overload and then the temperature switches inside of the transformer are opened Temperature switches are automat...

Страница 19: ...work includes parts and labour at Finero s cost but excludes shipping costs Finero products returned for warranty repair and found to be in proper working order may be subject to a shipping and handli...

Страница 20: ...ue to continuous development of the products warranty does not include updating of features When the warranty repair is booked the customer is requested to inform type and serial number of the equipme...

Страница 21: ...ver1 doc 20 27 Finero Oy The Quality Control Company Tel 358 5 7522 200 Haltonintie 5 FIN 47400 KAUSALA FINLAND Fax 358 5 3638 326 www qacontrol com sales qacontrol com ANNEX A HIGH VOLTAGE POWER FST...

Страница 22: ...h limit 100 990 10mA 990mA RAMPtype 0 Voltage increase or decrease type 1 up 2 up and down RLENght 5 Voltage increase or decrease time 0 4 30sec ON Test voltage on OFF Test voltage off ALARm ON Tester...

Страница 23: ...00 Test voltage 1000V VOLT 240 Test voltage 240V CURRent MAXimum Setting current high limit value 10 990 10 990mA mmm current high limit If measurement value exceed high limit voltage increasing stops...

Страница 24: ...ome status bits see STAT Example DCP VOLT 500 Commands to power test voltage 500V CURR MAX 100 current high limit 100 mA test ON voltage on and measurements starts OFF Test voltage off same like Stop...

Страница 25: ...Tester identify code free setting maximum 20 characters not marks Normal there are tester type and serial number Example IDN FST DCP sno 12345 RES Give test results leakage current test voltage statu...

Страница 26: ...ge return answer in real voltage in test 0 2000 Example DCP VOLT 500 Commands to power test voltage 500V CURR MAX 100 current high limit 100mA test ON voltage on and measurements starts VOLT Test volt...

Страница 27: ...o power test voltage 500V CURR MAX 100 current high limit 100mA test ON voltage on and measurements starts STAT Tester status inquiry If status byte 1 are 1 or 2 test are done If status byte 1 are 1 t...

Страница 28: ...LAND Fax 358 5 3638 326 www qacontrol com sales qacontrol com bb Input 2 status 0 off 1 on CAL Tester adjustement date Return answer dd mm yyyy 27 08 2001 DCP VER Tester inside circuits pcb and softwa...

Отзывы: