43
42
DC Current
AC Current (40 - 1 kHz)
Burden Voltage : 0.03 V/A
Ohms
Open Circuit Voltage : 1.3 Vdc
* Using Relative mode
Diode Tester
ms-Pulse/Duty Cycle
: Meas ure t he ti me signal is ON or O F F i n
milliseconds or in %
Close-Case calibration
: No internal adjustments needed
Battery/Fuse Access Door
: B attery or f use replaceable without voiding
calibration
High-Impact Overmolded Case
: Protective holster features
Electrical Specifications
Accuracy is given as ± ([% of reading] + [number of digits]), or otherwise specified,
at 23 C ± 5 C and less than 80 % R.H. for a period of one year after calibration.
DC Voltage
NMRR
: > 60 dB @ 50/60 Hz
CMRR
: > 120 dB @ DC 50/60 Hz, Rs=1 k
Input Impedance
: 10 M , 30 pF nominal
(50 M , 100 pF nominal for 500 mV range)
AC Voltage
CMRR
: > 60 dB @ DC to 60 Hz, RS = 1 k
Input Impedance
: 10 M , 30 pF nominal
(50 M , 100 pF nominal for 500 mV range)
Range
Accuracy
500.00 mV
0.01 mV
5.0000 V
0.1 mV
0.1 % + 2 d
50.000 V
1 mV
500.00 V
0.01 V
1000 V
0.1 V
Resolution
Range
Accuracy
500.00 mV
0.01 mV
1.0 % + 10 d
5.0000 V
0.1 mV
50.000 V
1 mV
500.00 V
0.01 V
Resolution
1000.0 V
0.1 V
400 Hz - 2 kHz
0.5 % + 10 d
40 - 400 Hz
Range
Accuracy
5.0000 A
100 µA
10.000 A
1 mA
0.5 % + 10 d
0.5 % + 20 d
Resolution
Range
Accuracy
5.0000 A
100 µA
10.000 A
1 mA
0.75 % + 10 d
1.0 % + 20 d
Resolution
Range
Accuracy
500.00 *
0.01
5.0000 k
0.1
50.000 k
1
500.00 k
0.01 k
5.0000 M
0.1 k
Resolution
50.00 M
10 k
0.1 % + 5 d
0.1 % + 2 d
0.1 % + 2 d
0.1 % + 2 d
0.3 % + 5 d
0.75 % + 10 d
Range
Open Circuit Voltage
2.0000 V
2.0 % + 1 d
< 3.0 Vdc
Accuracy