54
ad value of LD1,
weight value of LD2
and ad value of LD2
d2
= weight value of LD1. Data length = 8 including decimal
with leading space (Hex code 20).
d3
= ad value of LD1. Data length = 8 (integers only) with
leading space (Hex code 20).
d4
= weight value of LD2. Data length = 8 including decimal
with leading space (Hex code 20).
d5
= ad value of LD2. Data length = 8 (integers only) with
leading space (Hex code 20).
Aj
ad value of zero point
(offset), span weight
value, net span AD
value
d1
= as value of zero point. Data length = 8 (integers only) with
leading space (Hex code 20).
d2
= span weight value. Data length =8 including decimal with
leading space (Hex code 20).
d3
= net span AD value. Data length = Data length = 8 (integers
only) with leading space (Hex code 20)
Ak
Weight unit
enable/disable for kg,
g and lb
d1
= kg: - 0 = Off; 1 = On
d2
= g: - 0 = Off; 1 = On
d3
= lb: - 0 = Off; 1 = On
Al
Filter strength
0 = level 1; 1 = level 2; … ;7 = level 8; 8 = level 9
Am
Initial Zero range,
manual zero range,
auto zero tracking
speed
d1
= initial zero range: - 0 = Off, 1 = 1%, 2 = 2%; … 5 = 5%; 6 =
10%; 7 = 20%
d2
= manual zero range: - 0 = 1%; 1 = 2%; 2 = 3%; 3 = 4%; 4 =
5%; 5 = 10%; 6 = 20%; 7 = 50%; 8 = 75%; 9 = 100%
d3
= auto zero tracking speed: - 00 = Off; 01 = 0.25e; 02 =
0.50e; 03 = 10e; 04 = 1.5e; 05 = 2.0e; 06 = 2.5e; 07 = 3.0e; ….;
09 = 5e; 10 = 7.5e; 11 = 10e
An
Auto tare, repetitive
tare and preset tare
d1
= auto tare: - 0 = Off; 1 = On; 2= Continuous
d2
= repetitive tare: - 0 = Off; 1 = On
d3
= Preset tare: - 0 = Off; 1 = On
Ao
Stability control of
manual zero and
manual tare
d1
= manual zero stability control: - 0 = no (disable); 1 = Yes
(enable)
d2
= manual tare stability control: - 0 = no (disable); 1 = Yes
(enable)
Ap
Auto power off time,
backlight brightness,
auto power saving,
keypad buzzer and
system buzzer
d1
= auto power off time: - 0 = Off; 1 = 1 minute; 2 = 3 minute; 3
= 5 minute; 4 = 10 minute; 5 = 20 minute
d2
= backlight brightness (01 ~ 04). 01 = lowest, 04 = highest.
d3
= auto power saving: - 00 = Off; 01 = On
d4
= keypad buzzer: - 0 = Off; 1 = On
d5
= system buzzer: - 0 = Off; 1 = On
Aq
Check result buzzer,
near zero value, high
limit for weighing, low
limit for weighing,
high limit for
counting, low limit for
counting
d1
= check result buzzer: - 0 = Off; 1 = In; 2 = Out; 3 = hi, 4 = lo
d2
= near zero value. Data length = 8 with leading space (Hex
code 20)
d3
= Hi Limit for weighing. Integer only. Data length = 8
including decimal with leading space (Hex code 20).
d4
= Lo Limit for weighing. Integer only. Data length = 8
including decimal with leading space (Hex code 20).
d5:
= Hi Limit for counting. Integer only. Data length = 8
(integers only) with leading space (Hex code 20).
d6
: = Lo Limit for counting. Integer only. Data length = 8
(integers only) with leading space (Hex code 20).
Ar
Comport 1 working
mode, baud rate,
output protocol, print
d1
= working mode: - 0 = Auto 1; 1 = Auto 2; 2 = Auto 3; 3 =
Manual; 4 = PC; 5 = Scanner; 6 = Off
Содержание 19eA
Страница 2: ......
Страница 47: ...47 Appendix B1 Predefined PC Output Protocols ...
Страница 48: ...48 ...
Страница 49: ...49 ...
Страница 62: ...62 Fidelity Measurement Co Ltd www fi measurement com e mail info fi measurement com ...