5.4 Internal Function Table
Refer to the below tables for internal function number, parameter and setting notes.
No.
Description
Parameters / Note
Default = **
F1
Internal Analogue to Digital (ad) Value
Press [Print/M+] to set offset value to zero when unloaded. Then add load on the platform to observe the span value of load applied.
When ADC is more than 1 million, LO Light Bar will appear. Actual ADC is = 1 million plus the ADC value being displayed.
When ADC is more than 2 million, OK Light Bar will appear. Actual ADC is = 2 million plus the ADC value being displayed.
When ADC is more than 3 million, HI Light Bar will appear. Actual ADC is = 3 million plus the ADC value being displayed.
Press [Zero] to quit to F1.
F2
All Segment & Battery Voltage Check
All display segments and capacity bars will be lit on. Check any segments or LED of bars are missing.
To check battery voltage, press
[Check]
.
F3
Capacity, Division & Default Weight Unit
Display basic metrology characteristics (capacity, division and weight unit). Value displayed = Max + 1e
F4
Date Format & Date
DD/MM/YY
** YY/MM/DD
MM/DD/YY
Press [Print/M+] to check current date value. To change date value, press [Print/M+] then enter date value and then press [Print/M+] to confirm.
F5
Time
HH/MM/SS
To change time, press [Print/M+], then enter a new value and press [Print/M+].
F6
System Initialization
(Set F7~F32) to Default)
** NO
YES
If YES is selected, press [1] when “SURE?” is displayed. Indicator shows Done when initialization is completed.
F7
Auto Power Off Time (Minute)
OFF
1
3
** 5
10
20
Auto power off function will be disabled when an energized power adaptor is plugged in.
F8
Brightness
1
2
3
** 4
5
1 = dimmest; 5 = brightest
Press
[Func]
or
[Unit]
to change setting and then press
[Print/M+]
.
F9
Weight Unit Enable / Disable
kg
(
** On
/Off)
g
(On/
** Off
)
lb
(On/
** Off
)
F10
Filter Strength
1
2
3
4
** 5
6
7
8
9
Select: -
1 (strongest filter) for bad working environment where vibration, wind flow… etc. affect stable reading,
5 for normal environment,
9 (least filter) for very good working environment where wind and vibration have no effect to stable reading.
Press
[Zero
] to quit, or press
[Set]
to select conversion speed.
4 parameters are available.
Press
[Func]
or
[Unit]
to select: -