43
10. Peak Hold Mode
32
10.1 Description of Peak Hold Mode
Under this mode, the instrument will display and hold the highest load/force
detected. This mode can be used for tension (positive) or compression
(negative) tests, all peak results are treated as absolute values.
10.2 Comport Settings for Peak Hold Mode
33
Refer to
5.5
on comport settings. Output type parameters Auto 1 ~ 3 are
not
suggested. If a printer is used, set the connected comport to Manual and the
other comport to any data string related modes.
During comport setting procedures, following the below recommendation for
parameters selection.
a.
Baud rate setting. If an external peripheral is used, always use the
highest available baud rate of it. The highest baud rate this
instrument can support is 115200. Baud rate of the peripheral has
to be set accordingly.
b.
Parity setting: -
Set according to the external peripheral connected.
Select any if there is no external peripheral is connected.
c.
Data length setting
Set according to the external peripheral connected.
Select any if there is no external peripheral is connected.
d.
Auto Accumulation setting: - Select any. Instrument will deem No
as default value for this mode.
e.
Check Control setting: - Select Off.
f.
Stability control setting: - Select any. Instrument will deem No as
default value for this mode.
g.
Minimum output setting: - Select any. Instrument will deem 0d as
default value for this mode.
h.
Print format setting: - Set according to the external peripheral
32
Peak hold mode does not support memory accumulation, weight unit conversation or
weight check function.
33
Set both Comport 1 and Comport 2 to Off if no external peripheral are connected to this
instrument.