18
5.4 Internal Function Table
Refer to the below tables for internal function number, parameter and setting
notes.
No.
Description
Parameters / Note
Default = **
F1
Internal
Analogue to
Digital (ad)
Value
Press [Print/M+] to set offset value to zero when unloaded.
Then add load on the platform to observe the span value of
load applied.
When ADC is more than 1 million. Light Bar 1 will
appear. Actual ADC is = 1 million plus the ADC
value being displayed.
When ADC is more than 2 million. Light Bar 2 will
appear. Actual ADC is = 2 million plus the ADC
value being displayed.
When ADC is more than 3 million. Light Bar 3 will
appear. Actual ADC is = 3 million plus the ADC
value being displayed.
Press [Zero] to quit to F1.
F2
All Segment
Check
All display segments and capacity bars will be lit on. Check
any segments or LED of bars are missing.
F3
Capacity,
Division &
Default
Weight Unit
Display basic metrology characteristics (capacity, division
and weight unit). Value displayed = Max + 1e
F4
Date Format
& Date
DD/MM/YY
** YY/MM/DD
MM/DD/YY
Press [Print/M+] to check current date value. To change date value, enter
date value and then press [Print/M+] to confirm.
F5
Time
HH/MM/SS
To change time, press [Print/M+], then enter a new value and press [Print/M+].
F6
System
Initialization
(Set F7~F31
to Default)
** NO
YES