19
If YES is selected, press [1] when “SURE ?” is displayed. Indicator shows
Done when initialization is completed.
F7
Auto Power
Off Time
(Minute)
OFF
1
3
** 5
10
20
Auto power off function will be disabled when an energized power adaptor is
plugged in.
F8
Brightness
1
2
** 3
4
1 = dimmest; 4 = brightest
Press
[Func]
or
[Unit]
to change setting and then press
[Print/M+]
.
F9
Weight Unit
Enable /
Disable
kg
(
** On
/Off)
g
(On/
** Off
)
lb
(On/
** Off
)
F10
Filter
Strength
1
2
3
4
** 5
6
7
8
9
Select: -
1 (strongest filter) for bad working environment where vibration,
wind flow… etc. affect stable reading,
5 for normal environment,
9 (least filter) for very good working environment where wind and
vibration have no effect to stable reading.
F11
Auxiliary
Function
Cnt (Counting) / On/
** Off
AtM (Action Tare Memory) / On/
** Off
PEK (Peak) / On/
** Off
Ani (Animal) / On/
** Off
PCd (Quick Access to Product Code Setting) /
On/
** Off
F12
Auto Tare
Function
** OFF
ON
Contin
Notes: -
Off = Auto Tare Function disable.
On = Only the first table weight applied will be tare off. Minimum
tare load ≥ 2d.