USER INSTRUCTIONS
Cleaning and maintenance
Cleaning
Clean the heaterbase, sensor cartridge, or expiratory heater wire adapter using a cloth dampened with either isopropyl
alcohol or neutral detergent� Always disconnect the humidifier from the power supply before cleaning�
NOTES:
• Do not immerse or autoclave the heaterbase, sensor cartridge, or expiratory heater wire adapter�
• Do not spray liquid into the vents or onto electrical connectors� Failure to comply may result in irreparable damage to the
humidifier�
Routine maintenance
A full technical description, including routine maintenance and service data, is contained in the Product Technical Manual
available from your supplier or Fisher & Paykel Healthcare�
WARNING:
The Product Technical Manual must be followed for all servicing and maintenance of the
humidifier� Failure to comply may impair performance of the humidifier or compromise safety (including potentially
causing serious harm)�
Warnings, cautions and notes
WARNINGS
•
Refer to the instructions for use for breathing circuits, interfaces and accessories before operating the equipment�
Failure to comply may impair performance of the humidifier or compromise safety (including potentially causing patient
harm)�
•
This product is only designed and verified for use with accessories and spare parts approved by Fisher & Paykel
Healthcare� Unauthorized accessories or spare parts which are used with the humidifier may impair performance
of the humidifier, or compromise safety (including potentially causing serious patient harm), or result in increased
electromagnetic emissions, or decreased electromagnetic immunity, resulting in improper operation�
• This product is designed for the delivery of air and/or oxygen� It is not suitable for the delivery of flammable anesthetic
gas mixes or Heliox gas� Failure to comply may impair performance of the humidifier or compromise safety (including
potentially causing patient harm)�
• The humidifier should always be level and positioned lower than the patient� Failure to comply may impair performance
of the humidifier or compromise safety (including potentially causing serious patient harm)�
• Visually inspect components and accessories for damage before use and replace if damaged� Use of damaged
components or accessories may impair performance of the humidifier or compromise safety (including potentially
causing serious harm)�
• Appropriate patient monitoring (e�g� oxygen saturation) must be used at all times� Failure to monitor the patient (e�g� in
the event of an interruption to gas flow) may result in serious harm or death�
• Do not touch the electrical connectors and the patient simultaneously� Failure to comply may result in serious harm�
• Operation of the humidifier outside of the recommended operating conditions (as described in these user instructions)
may impair performance of the humidifier or compromise safety (including potentially causing patient harm)�
• Monitor circuit condensate every six hours to prevent occlusion or build-up of fluid� Drain as required� Failure to comply
may impair performance of the humidifier or compromise safety (including potentially causing serious patient harm)�
• Remove any sources of ignition, such as: cigarettes, an open flame, or materials which ignite easily at high oxygen
concentrations�
• Follow the instructions of the oxygen device provider; keep oxygen regulators, cylinder valves, tubing, connections, and
all other oxygen equipment away from oil, grease, or greasy substances� Spontaneous and violent ignition may occur if
these substances come into contact with oxygen under pressure�