9
USER INSTRUCTIONS
User interface
Comfort settings
With an adult or pediatric inspiratory limb connected, it is possible to change the target temperature in Mask and
Optiflow modes, to provide conditions which may encourage patient comfort�
When additional neonatal modes are enabled, changing the target temperature in CPAP | NIV and Optiflow modes is also
possible�
1
3
2
4
The available comfort settings are:
Adult & Pediatric
Neonatal
Mode
Default
Medium
Low
Mode
Default
Medium
Low
Invasive
37 °C
-
-
Neonatal
37 °C
-
-
Mask
31 °C
29 °C
27 °C
Invasive*
37 °C
-
-
Optiflow
37 °C
35 °C
33 °C
CPAP | NIV*
37 °C
34 °C
31 °C
Optiflow*
37 °C
35 °C
33 °C
* with additional modes enabled
The humidifier will reset to the default set-point if the mode is changed or the humidifier is turned off and back on� It is
possible for service personnel to change the default set-point for Mask, CPAP | NIV and Optiflow modes in the service
menu�