USER INSTRUCTIONS
Technical specifications
Product specifications
Heaterbase Specifications
Dimensions (heaterbase
only)
240 mm (D) x 154 mm (W) x 253 mm (H)
Weight (heaterbase and
power cord only)
3�45 kg
Supply frequency
50/60 Hz
Supply voltage
950AXX
1
230 V
950JXX
1
115 V
950GXX
1
100 V
Power rating
350 VA
Maximum length of power
cord
3�3 m
Sound pressure level
Alarms exceed 45 dbA @ 1 m
Auditory alarm pause
120 seconds
Maximum temperature of
delivered gas
43 °C
Time to reach set
temperature (gas flow is
required)
< 30 minutes
Maximum surface
temperature of the breathing
circuit (applied part section)
44 °C
Component service life
Heaterbase: 7 years
Adult
Pediatric
Neonatal
Humidity performance
(Except in the event of a
humidifier alarm or power
failure or electromagnetic
disturbance)
Invasive mode:
> 33 mg/L
Mask mode:
> 12 mg/L
Optiflow mode:
> 12 mg/L
Invasive mode:
> 33 mg/L
Mask mode:
> 12 mg/L
Optiflow mode:
> 12 mg/L
Neonatal Mode:
> 33 mg/L
Invasive mode:
> 33 mg/L
CPAP | NIV mode:
> 12 mg/L
Optiflow mode:
> 12 mg/L
Operating flow range
(L/min, STPD)
Invasive mode:
5-60 L/min
Mask mode:
5-120 L/min
Optiflow mode:
5-70 L/min
Invasive mode:
1-60 L/min
Mask mode:
1-60 L/min
Optiflow mode:
1-60 L/min
Neonatal Mode:
0�5-40 L/min
Invasive mode:
0�5-40 L/min
CPAP | NIV mode:
0�5-40 L/min
Optiflow mode:
0�5-36 L/min
1
XX represents the country code