![F.W. Bell 9900 Series Скачать руководство пользователя страница 50](http://html.mh-extra.com/html/f-w-bell/9900-series/9900-series_operation-and-maintenance-manual_535142050.webp)
SERIES
9900
9000
SERIES
GAUSSMETERS
SECTION VII continued
V
I
I
-C DIAGNOSTIC ERROR CODES
Diagnostic Error Codes
The 5-digit
major
and
minor
error codes help to pinpoint a failure within the
various sections of the gaussmeter. The first digit indicates which subsection
failed, with the remaining four digits providing precise information about the
failure. Typically, only the first one or two digits are of any practical value to the
user. The remaining information provides F.W. Bell with component-level
information. The user should not attempt to make component-level repairs.
Field repairs should be limited to subassembly replacements
only.
Returned
assemblies should have with them the error code and a brief description of
the problem.
VII-C(a) ERROR Oxxxx
This MAJOR error indicates a failure on the
processor card (item
#201500).
VII-C(b) ERROR 1xxxx
This MAJOR error indicates a failure of the
peripheral control card
(item
#201501).
VII-C(c) ERROR 2xxxx
This MAJOR error indicates a failure on the SERIES-9900
power supply
controller card (item #201505) or the line voltage input transformer assembly
(item #201507).
NOTE: A failure in one of the probe interface modules can cause a failure of
the power supply. Before attempting repairs on the power supply, try a
systematic replacement (or removal) of each probe interface module.
VII-C(d) ERROR 31xxx
This MINOR error indicates a failure of the
processor card (item
#201500) that
prevents operation of the CHANNEL 1 (left-most) probe interface module.
Operation can continue with CHANNELS 2 and 3, but the
processor card
will
have to be replaced to operate with all three channels.
VII-C(e) ERROR 32xxx
This MINOR error indicates a failure on the
processor card (item
#201500)
that prevents operation of the CHANNEL 2 (middle) probe interface module.
Operation can continue with CHANNELS 1 and 3, but the
processor card
will
have to be replaced to operate with all three channels.
VII-C(f) ERROR 33xxx
This
MINOR
error indicates a failure on the
processor card (item
#201500)
that prevents operation of the CHANNEL 3 (right-most) probe interface
module. Operation can continue with CHANNELS 1 and 2, but the
processor
card
will have to be replaced to operate with all three channels.
VII-C(g) ERROR 34xxx
This MINOR error indicates a failure on the
processor card (item
#201500)
that prevents operation of the RS-232 communications port.
45